USE SCENARIO EXAMPLE: Assessment of the manufacturer’s rationale for providing information physically or digitally for a medical device handled by lay users
STORYLINE: A manufacturer develops a medical device intended to be handled and used by lay users and implements a labelling concept in which information is distributed between the device, its packaging and a digital label.
As part of product development, the manufacturer considers the actual handling and use of the device by the intended lay-user population and identifies the information needed at different stages, for example:
Select device → check suitability → prepare device → use device → respond to problems → store/dispose of device
Through its risk management and usability engineering processes, the manufacturer evaluates when specific information is needed, what could happen if the information is unavailable or misunderstood, and whether digital access is appropriate in the relevant use situation. Based on this assessment, the manufacturer determines which information is provided physically, digitally or through both routes. For example, the manufacturer may conclude that information such as an expiry date, sterile status or information required to select the correct product variant needs to be immediately available, while detailed troubleshooting information can be provided digitally.
During conformity assessment, the Notified Body reviews the manufacturer’s documentation to determine whether this approach is plausible and adequately supported by the manufacturer’s risk management, usability engineering and other relevant processes. The NB particularly considers whether the manufacturer has adequately addressed foreseeable situations in which a lay user handling the device cannot access digitally provided information at the time it is needed.
PRIMARY PERSONA AND SUPPORTING PERSONAS WHERE RELEVANT: Notified Body Technical Documentation Reviewer/Auditor; Supporting Personas; Notified Body usability and/or risk-management specialist; Manufacturer Regulatory Affairs / Labelling representative; Manufacturer Risk Management representative; Manufacturer Usability Engineering representative; Representative lay users involved in usability evaluation/validation
INFORMATION NEEDS: The NB needs sufficient information in the manufacturer’s Technical Documentation to understand how and why the manufacturer arrived at its labelling concept.
Intended user and use context
The NB needs information on: intended lay-user population; relevant user characteristics and foreseeable limitations; intended use environment; relevant steps involved in handling and using the device; circumstances in which information is needed.
Information needs identified by the manufacturer
The documentation should enable the NB to understand:
What information does the lay user need at each relevant stage of handling/use?
When is the information needed?
What user decision or action depends on that information?
What are the consequences if the information is unavailable, overlooked or misunderstood?
This could concern, for example: product identification; product/variant selection; expiry date; sterile status and package integrity; storage conditions; preparation for use; essential warnings or precautions; troubleshooting; post-use handling or disposal.
Manufacturer’s rationale for the information route
The NB needs to understand which information the manufacturer has decided to provide: on the device; on the packaging; digitally; through more than one route. For information provided exclusively digitally, the documentation should demonstrate why the manufacturer considers digital access appropriate for the identified information need and use situation.
Supporting evidence
The rationale should be traceable, as applicable, to the manufacturer’s: Risk Management File; Usability Engineering File; information-for-safety decisions; labelling specifications; evaluation of foreseeable misuse; usability verification/validation; assessment of foreseeable digital-access failures; physical and digital label specifications.
The core information need for the NB is therefore: Can the rationale for the information provided physically and digitally be understood and substantiated from the manufacturer’s documentation?
ROUTE(S) OF ACCESS TO INFORMATION: Today: The NB typically obtains the relevant evidence from different parts of the Technical Documentation:
Device description → Risk Management File → Usability Engineering File → label/packaging → IFU → verification and validation evidence
The reviewer determines whether the resulting information supplied with the device is consistent with the manufacturer’s identified risks, intended users and conditions of use.
Target: With digital labelling, the assessment additionally needs to capture the manufacturer’s rationale for the information architecture across different channels:
Risk Management + Usability Engineering + user information needs → manufacturer’s labelling decision → physical information / digital information / both → verification and validation
The NB should also be able to understand the resulting user journey:
Lay user handles device → information need arises → information source is available → user accesses and understands information → appropriate action can be taken
Importantly, the NB is assessing the manufacturer’s process and resulting justification, rather than designing this user journey itself.
TRUSTWORTHINESS/QUALITY CONSIDERATIONS: Manufacturers QMS; Manufacturer’s technical documentation
PRELIMINARY KPI CONSIDERATIONS: N/A for regulatory use scenario
ADDITIONAL COMMENTS: Use scenario is more related to information needed by the NB during conformity assessment (audit/technical documentation)